The solution to the tilt of the electric operating table
2019-05-22
The electric operating table is an increasingly popular device. It can freely rotate and convert angles, which can greatly reduce the labor of the staff. It is very suitable for urinary, gynecological, and orthopedic operations. But after using for a long time, you may find that the electric operating table is tilted. What is the cause of this, and how to solve it? Let’s understand together:
First, let's judge whether the solenoid valve is faulty. There are two ways to judge: one is to measure the resistance with a three-meter, and the other is to use metal to see if there is suction. Of course, if the solenoid valve pulls in action is normal. The above-mentioned problems can also occur when the oil circuit is blocked.
Then determine if the compression pump is malfunctioning. First check whether the voltage on the compression pump is present, and use a three-meter to measure the resistance of the compression pump. If the foregoing is normal, the commutation capacitor is generally invalid. The electric operating table has movement in one direction and no movement in the other direction. The unilateral non-action failure is generally caused by the electromagnetic directional valve. The electromagnetic reversing valve failure may be caused by a bad control circuit, or the reversing valve may be mechanically stuck. The specific inspection method is to first measure whether the directional valve has voltage. If there is voltage on the operating bed, disassemble and clean the reversing valve.
Due to long-term use, if any small impurities on the movable shaft of the interrogation valve will pull the shaft and cause it to jam, the operating table will only move in one direction. When the operating table is in use, it will automatically lower, but the speed is very slow. This situation occurs more frequently in mechanical operating tables, mainly due to failure of the lift pump. After the operating table is used for several years, very small impurities may stay in the oil inlet valve port, causing a small internal leakage. The solution is to disassemble the lift pump and clean it with gasoline, especially to check the oil inlet valve.
The above is the solution to the tilt of the electric operating table. If you want to know more information, please contact us.
Related News
2026-08-17
Maximize Surgical Efficiency with Integral Reflection Type Surgical Shadowless Lamp
2026-08-06
Operation Shadowless Lamp: Revolutionizing Surgical Lighting for Optimal Precision
2026-07-30
Exploring the Full Netcom Intelligent Digital Operating Room: A Leap Forward in Surgical Technology
2026-07-23
Unveiling the Versatility of Modern Multifunctional Operating Tables: A Comprehensive Guide
2026-07-16
Enhancing Surgical Precision: The Role of Light-Operating Ceilings in Modern Operating Rooms
2026-07-09
Guiding the Surgeon’s Hand: The Evolution of Portable Operating Lights
Esperanto
Shqiptare
Euskara
Zulu
Latinus
Cymraeg
தமிழ்
Slovak
Slovak
Afrikaans
CONTACT US
Foreign trade mailbox: sales5@jhsurgitech.com
Foreign trade mailbox: ervin@jhsurgitech.com
Foreign trade mailbox: sales@jhsurgitech.com
Address: No.68, South District, Shigang Industrial Park,
Tongzhou District, Nantong City, Jiangsu Province
Leaving a message.
© 2022 NANTONG MEDICAL APPARATUS CO.,LTD. 苏ICP备20010639号 SEO